Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202049

    Validate with FDA (Live)

    Date Cleared
    2021-09-07

    (410 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Python catheters are indicated for removal of thromboemboli from the peripheral arterial system, and for occlusion and infusion of fluids into a vessel.
    The Over-the-Wire Latis Graft Cleaning Catheter is indicated for the removal of thromboemboli from vascular grafts, and for occlusion and infusion of fluids into a graft.

    Device Description

    The Python Embolectomy Catheter & Over-the-Wire Latis Graft Cleaning Catheters are sterile, single use vascular balloon catheters intended for use in the removal of thromboemboli, occlusion of vessels, and the infusion of fluids. The devices are intended for use in arterial vessels and artificial vascular grafts respectively. The body of the catheters utilize a dual lumen design with one inflation lumen used to inflate the balloon, and one through lumen allowing for insertion over a guide wire or infusion of fluids. The outer diameter of the body is designed to accommodate use with a standard percutaneous access sheath. The Over-the-wire Latis Graft Cleaning Catheter includes a braided mesh which surrounds the balloon and is designed to increase engagement with wall of artificial vascular graft.

    AI/ML Overview

    The provided text describes a medical device's 510(k) submission to the FDA, demonstrating substantial equivalence to predicate devices. It focuses on the physical and material characteristics of the catheters and the bench testing performed. However, it does not contain information pertinent to the typical acceptance criteria or study design for AI/ML-based medical devices, such as a test set size, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set details.

    Therefore, I am unable to provide the requested information based on the input text.

    The document discusses the following:

    • Device Name: Python Catheter and Over-the-Wire Latis Graft Cleaning Catheter
    • Intended Use: Removal of thromboemboli from the peripheral arterial system/vascular grafts, and for occlusion and infusion of fluids into a vessel/graft.
    • Testing Performed:
      • Engineering/Bench Testing: Balloon Pull Force, Inflation & Guidewire Hub Pull Force, Catheter Body Tensile Strength, Catheter Balloon Cycling Test, Balloon Burst Pressure Test, Inflated Balloon Diameter Inspection, Guidewire & Inflation Hub Leak Test, Simulated Use Bend Test, Torsion Test, Kink Test.
      • Biocompatibility Testing: Cytotoxicity, Intracutaneous Irritation, Sensitization, Acute Systemic Toxicity, Hemolysis, Materials Mediated Pyrogenicity, Platelet and Leukocyte Counts, Partial Thromboplastin Time, Complement Activation, Thrombogenicity.
    • Standards Referenced: ISO 10555-1:ed2:2017 and EN 13868.

    This information relates to the physical and biological safety and performance of the catheter itself, not to the performance of an AI/ML component.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1